Fluke Biomedical Cufflink User Manual

Page 83

Advertising
background image

Remote Operation

Command Descriptions

3

3-9

Parameters for this command are described below.

Parameter Definition

Range Default Description

HR Heart

Rate

(Beats/Minute)

30 BPM to 240
BPM in 1 BPM
Steps

80 BPM

1 BPM resolution

BP Blood

Pressure

(mmHg)

ADAMS-60/30

ADAMS-80/50

ADAMS-100/65

ADAMS-120/80

ADAMS-150/100

ADAMS-200/150

ADAMS-255/195

ADAMS-
120/80

PULSEID Pulse

Identification
(milliseconds)

ID# / WIDTH

0 / 800 ms

1 / 500 ms

2 / 250 ms

3 / 720 ms

4 / 230 ms

5 / 280 ms

6 / 350 ms

7 / 480 ms

8 / 980 ms

9 / 1980 ms

10 / 1480 ms

ID# 3
(720 ms)

Pulse width affects the rise
time of pulse

ENVGAIN Pressure

Envelope Gain

1 to 200%
(10 to 2000
RVDUs)

100% is a
typical value
for adult
cuffs.

Affects pulse strength or
amplitude

ENVSHIFT Pressure

Envelope Shift

-100 to +100
mmHg

0 mmHg

Modifies existing pressure
envelope to simulate another
BP

Example: 0 +10 mmHg shift
causes 120/80 to read
130/90 on most monitors.

Example

Draw a pressure envelope showing a BP of 80/50 with a heart rate of 60 BPM

Command

drawenv

hx=60,bp=adams-80/50

Results

The specified pressure envelope is drawn on the
Analyzer display: