Cufflink

Operators Manual

3-12

Parameters for this command are described below.

Parameter Definition

Range Default

Description

HR Heart

Rate

(Beats/Minute)

30 BPM to 240 BPM
in 1 BPM Steps

80 BPM

1 BPM resolution

BP Blood

Pressure

(mmHg)

ADAMSNEO-60/30
ADAMSNEO-80/50
ADAMSNEO-100/65
ADAMSNEO-120/80
ADAMSNEO-
150/100

ADAMSNEO-
120/80

PULSEID Pulse

Identification
(milliseconds)

ID# / WIDTH

0 / 800 ms
1 / 500 ms
2 / 250 ms
3 / 720 ms
4 / 230 ms
5 / 280 ms
6 / 350 ms
7 / 480 ms
8 / 980 ms
9 / 1980 ms
10 / 1480 ms

ID# 3
(720 ms)

Pulse width affects
the rise time of
pulse

ENVGAIN Pressure

Envelope Gain

1 to 200%
(10 to 2000 RVDUs)

100% is a
typical value
for neonate
cuffs.

Affects pulse
strength or
amplitude

ENVSHIFT Pressure

Envelope Shift

-100 to +100 mmHg

0 mmHg

Modifies existing
pressure envelope
to simulate
another BP

Example: 0 +10
mmHg shift
causes 120/80 to
read 130/90 on
most monitors.