Utility functions, Set clock – Fluke Biomedical Cufflink User Manual

Page 50

Advertising
background image

Cufflink

Operators Manual

2-22

4. Press

F2 Reset

to return

Pressure actual

and

Pressure peak

to zero. If

Pressure actual

is still not zero, press

F5 Zero Pressure

.

5. Press

F3 Print

to print the test results. Table 2-5 describes the measured parameters.

Table 2-7. Measured Pop Off Test Parameters

Parameter Description

Pressure Actual (mmHg)

The current pressure in the blood pressure cuff

Pressure Peak (mmHg)

The maximum pressure in the system before
release of the monitor's emergency valve

6. Press

F4 Release Pressure

to release all pressure in the system.

7. Press the

Esc

key to quit the

Manometer

mode.

Utility Functions

The analyzer enables the user to set the clock, adjust pop time, view the logo display, and

perform system tests from the

Util

submenu:

F1

F2

F3

F4

F5

Select BP Press Util

Util Comm Auto

Set Clock

Set Clock
Pop Time
Logo
System

Set calendar clock

fcv098.eps

All of these functions are system tests or adjustments and affect only the Analyzer itself.

The NIBP monitor is not affected by the

Util

options.

Set Clock

The

Set Clock

function allows changes to be made to the clock or calendar, as shown in

the upper right corner of the logo display. The time is displayed in an hour, minutes,

seconds format, while the date is displayed in the date, day, year format.
To set the Analyzer clock:
1. From the

Util

submenu, highlight

Set Clock

and press the

Ent

key. The

Set Clock

( labeled

Set Time and Date

) screen displays: