Fill burette

1.

Enter the Burette functions menu:

2.

Select the burette in the Address field,

see "Address", page

38

.

3.

Press 1 Fill. The burette is filled at maximum speed using the
installed reagent. Approximate filling time: 25 s.

Flush burette

A flush rinses the burette, tubings and delivery tip thoroughly. It is
recommended to flush the burette on a daily basis.

The Titration Manager carries out a flush after each installation or rea-
gent replacement procedure.

To run a flush:

1.

Enter the Burette functions menu.

2.

Select the burette in the Address field (TitraLab 845 only).

3.

Press 3 Flush.

The flush procedure runs as follows:

Empty,

Fill 4 ml (*) then empty (5 times).

(*) 1 ml for a 10 ml burette and 10 ml for a 50 ml burette.

The length of the “Flush” procedure depends on the volume of the
burette.

You can also flush simultaneously all installed burettes of a
TitraLab 845,

see "Global flush burettes", page 103

.

Format
(printouts)

Format = Listing

The whole report is printed in one operation.

Format = Page by page

The printer waits until a preset number of lines have been collected
then prints one page (this number is set by the Nb line per
page parameter),

see "Nb lines per page (printouts)", page 124

.

The printing format applies for automatic printouts (at the end
of a test) or manual printouts (by pressing key Print).

Access:

Setup Menu > Configuration

Refer to "Printouts", page 136

.

Advertising