Start timer, Static rinses, Static rinse time – Hach-Lange TITRALAB 840_845 Reference Manual User Manual
Page 184: Statistics, Titra lab

Titra
Lab
®
TitraLab 840 and 845 Reference Manual
Page 184
Start timer
This is the time delay the Titration Manager will respect before starting
the titration. If a predose is programmed, this delay will be counted
down at the end of the predose. The Start timer delay will allow the
solution to stabilise before performing the first measurement.
While running the method, you can carry on regardless this
delay by pressing key Del.
Enter in:
Edit method > Parameters menu
Edit reagent > Calibration parameters menu
Range available:
00:00 to 99:59 min:s
Static rinses
Refer to "Number of static rinses", page 127
.
Static rinse time
Refer to "Rinse time", page 160
Statistics
When several tests are performed on the same sample, the mean and
standard deviations are calculated from the performed tests.