Start timer

This is the time delay the Titration Manager will respect before starting
the titration. If a predose is programmed, this delay will be counted
down at the end of the predose. The Start timer delay will allow the
solution to stabilise before performing the first measurement.

While running the method, you can carry on regardless this
delay by pressing key Del
.

Enter in:

Edit method > Parameters menu
Edit reagent > Calibration parameters menu

Range available:

00:00 to 99:59 min:s

Static rinses

Refer to "Number of static rinses", page 127

.

Static rinse time

Refer to "Rinse time", page 160

.

Statistics

When several tests are performed on the same sample, the mean and
standard deviations are calculated from the performed tests.