Proportional band, Titra lab – Hach-Lange TITRALAB 840_845 Reference Manual User Manual
Page 140

Titra
Lab
®
TitraLab 840 and 845 Reference Manual
Page 140
Proportional
band
This is the pH (mV) span prior to the end point, at which the titrant
speed will be reduced. The speed will increasingly reduce as the end
point is approached.
For titrations where the slope of the curve is high near the end point,
the proportional band setting should be large. If overshooting of the
end point frequently occurs, increase the Proportional band
or decrease the Max. speed setting. A high value for the Pro-
portional band will result in slow titrations.
Enter in:
Edit method > Parameters menu (end point method)
Edit reagent > Calibration parameters menu (end point method)
Range available:
-9.000 pH to +23.000 pH or -2000.0 mV to +2000.0 mV
End point no. n
Maximum
speed
Minimum
speed
Titrant delivery speed
mV or pH