Recalculate results, Titra lab – Hach-Lange TITRALAB 840_845 Reference Manual User Manual
Page 152
Titra
Lab
®
TitraLab 840 and 845 Reference Manual
Page 152
Recalculate
results
You can recalculate a result at the end of a sample titration or reagent
calibration method. The old result is replaced with the new. The
recalculated result is saved with the indication "Recalculated".
Proceed as follows to recalculate a result:
•
At end of titration, press 2 More details the 3 Recaculate in the
results screen.
•
Change the calculation parameters.
•
Recalculated results are displayed.
You can accept then save the recalculated results (press 1 Back then
1 Save and continue). To view the results,
The test quantity can be modified for each
test performed.
Press 2 to recalculate the result(s) or 1 to
return to the first screen without
recalculating the result(s).
You can start a new recalculation (press 2)
or return to the first screen above (press 1).