Solution menu, Titra lab – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 181

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 181

Solution menu

This menu is available for pH electrode that are edited with the
Calibrate request = Yes option.

To access:

1.

From the Electrode window, press 4.

2.

Select the electrode to be edited.

3.

Press 2 Edit electrode and check that the Calibration
request = Yes option has been selected.

4.

Use the LEFT/RIGHT arrow keys to move to the last
Edit electrode display.

5.

Press 2 Calibration solutions.

Select the pH standard solutions to be used for the calibration. The
pH values are given at 25°C. The following values are available:
IUPAC pH standards (pH 1.679, 4.005, 6.865, 7.000, 7.413, 9.180,
10.012 or 12.454 at 25°C), 4-7-10 Series (pH 4, 7 or 10 at 25°C) or
20°C series (pH 4.00 or 9.00 at 20°C).