Number of decimals, Number of digits, Number of dynamic rinses – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 125: Number of ep, Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 125

Number of
decimals

Number of decimals (0 to 3) to be displayed and printed for the result.

Enter in:

Edit method > Results (for a pH measurement method)

Number of
digits

Number of significant digits (1 to 5) to be displayed and printed for the
result calculated.

Example: If Number of digits = 4:

1456.1 is displayed “1456”

12.124 is displayed “12.12”

0.15872 is displayed “0.1587” (the first significant digit is “1”)

0.4 is displayed “0.4000” (the first significant digit is “4”)

Enter in:

Edit method > Results (for titration methods)
Edit reagent > Results

Number of
dynamic rinses

If a SAC850 Sample Changer is in use, enter the desired number of
dynamic rinses to be carried out before each beaker
analysis of a SAC sequence,

see also "Dynamic rinses", page 77

.

Enter in:

Setup menu > Configuration

Range available:

0 to 9

Number of EP

Number of end points (1 to 2) to be titrated to.

Only one end point is titrated to during a back titration or a
reagent calibration.

Enter in:

Edit method > Method parameters menu (for End point method)