Reagent unit, Reagent window, Titra lab – Hach-Lange TITRALAB 840_845 Reference Manual User Manual
Page 151
Advertising
Titra
Lab
®
TitraLab 840 and 845 Reference Manual
Page 151
Reagent unit
Specify the reagent concentration unit. The units are given on the
reagent bottle or are the unit used to express the titrant concentration
after calibration.
Enter in:
Reagent library > Edit reagent menu.
Range available:
Select, mM = mmol/l, M = mol/l, mN = meq/l or N = eq/l.
If molar units are selected (mM or M), it is necessary to enter
stoichiometric coefficients for the chemical reaction.
Once the units have been confirmed, they are added to the
reagent ID.
Reagent
window
This window contains all the information and operations concerning
the reagents.
To access:
Use LEFT/RIGHT arrow keys
Advertising