Balance in use

When using a balance connected to the Titration Manager, proceed
as follows:

1.

Press 1 to run the method.

2.

If required, enter a User ID, confirm.

3.

If required, enter a Sample ID.

4.

Place the sample in the sample container.

5.

Place the sample container on the balance and set the balance to
zero “tare”.

6.

Introduce the sample into the titration cell.

7.

Weigh the “empty” sample container

8.

Press

✓ in the field Test 1.

9.

Press

✓ to accept the sample amount.

10. Press 1 to continue the titration.

Make sure that the sample amount unit is expressed in g or
mg.

Bar code reader
connection

Connect a bar code reader to the Titration Manager via the 6-pin mini
DIN port situated on the right hand side of the instrument.