Printouts title, Programming method, Titra lab – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 138

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 138

Printouts title

Title of the report printout (1 to 23 characters).

Enter in:

Edit method > Printouts
Edit reagent > Printouts
Edit electrode > Printouts

Programming
method

1.

Check or create the electrode(s) to be used by the method.

Refer to "Create electrode", page 66

.

2.

Check or create the reagent(s) to be used by the method. If you
intend to perform a reagent calibration, make sure that the
electrode(s) used for calibration are the same as the one(s) used
in the method.

Refer to "Create reagent", page 67

.

3.

Finally, create the method, which will consequently use the
electrode(s) and reagent(s) created in the first two steps of pro-
gramming.

Refer to "Create method", page 66

.

Only the Supervisor is allowed to program the methods. Once
you have finished programming, make sure that NO question
marks "?" are displayed in the Reagent and Electrode tabs!

If “?” are displayed in the Reagent and Electrode tabs, press 1
in the Main window to check the method. The instrument indi-
cates the possible errors and prompts you to correct them, until
?” disappears.

Refer to "Programming methods", page 29

.