TitraLab 840 and 845 Reference Manual

Page 191

TIM cell external
Gnd

Specify in the Configuration menu if the grounding of the measuring
cell takes place using an external connection to the titration system.

This is the case when the solution is grounded via a metal shield or
via a conductivity cell connected to a conductivity meter.

The following configurations will be therefore not possible:

connecting a metal electrode to the GND socket of the titration
system,

connecting a double platinum electrode to the Pt-Pt socket of the
titration system.

If TIM cell external Gnd = Yes and a reference
electrode is connected to the Ref. socket of the titration system,
grounding will take place by an external link and not by the Ref.
socket.

If TIM cell external Gnd = No, grounding of the cell will be
determined in the method by the measurement type (pH, mV or mV
(i>0)) and the parameter Cell grounding (Reference/Metal/Other).

Refer to "Cell grounding", page 57

.

Time max
(result
indicator)

Refer to "Result indicators", page 157

.

Title

Enter the title of the calibration report sheet (max. 23 alphanumeric
characters).

Enter in:

Edit method > Printouts
Edit reagent > Printouts
Edit electrode > Printouts

Titrant ID

The name of the titrant used by the method.

The titrant can be selected from the Users list of available reagents or
created from the preprogrammed list of reagents.

If the titrant is calibrated, the electrode(s) used for the
calibration must be the same electrode(s) used in the method.

Enter in:

Edit method > Parameters menu