Reagent addition volume, Reagent calibration, Refer to "reagent – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 145: Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 145

Reagent
addition volume

Option available with TitraLab 845.

Volume of reagent to be added.

Enter in:

Edit method > Parameters (if Addition = Yes)
Edit reagent > Calibration parameters (if Addition = Yes)

Range available:

0.01 ml to 999 ml

Reagent
calibration

1.

Select the method which uses the reagent to be calibrated.

2.

If a Question mark "?" is present in the Reagent and/or Electrode
tabs, it means that the method needs to be programmed - a rea-
gent or an electrode is missing. Review programming in Supervi-
sor mode.

Refer to "Programming method", page 138

.

3.

Install the reagent(s),

see "Install reagent", page 108

.

4.

Connect the electrode,

see "Electrode connection", page 86

.

5.

Press 1 Calibrate/Enter titre in the Reagent window.

6.

Select the titrant from the list.

7.

Press 1 Run calibration and follow the messages on the display.

8.

Measurements start.

The titre ±standard deviation is calculated at the end of the
calibration cycle. The calibration results are saved with the
reagent.
To consult the calibration results,

see "GLP-Archives menu",

page 104

.

Standard ID.

Time elapsed since titration started

Temperature measured or entered

Potential measured

Titrant ID

Test number

Volume of titrant dispensed in ml

Titrant burette addition speed

Potential (mV) = f (volume dispensed (ml))

Advertising