The following tables give the conversion factor to use in the calculation equation.

If the result is in volume, the result given will be the volume Vi added to reach the equivalence
point. This volume will be converted in ml or µl if asked for.

R = V *c

with:

Sample volume Vs in ml:

Result

c (if ml conversion)

c (if µl conversion)

in ml by default

1

1000

“C

smp

“C

reag

“c” if “Vs” in “ml”

eq/l = N

eq/l = N

1

eq/l = N

meq/l

0.001

eq/l = N

mol/l = M

x

eq/l = N

mmol/l

0.001 x

meq/l

Same formula as C

smp

= eq/l but multiplied by 1000

mol/l = M

Same formula as C

smp

= eq/l but multiplied by 1 / y

mmol/l

Same formula as C

smp

= eq/l but multiplied by 1000 / y

g/l = mg/ml

Same formula as C

smp

= eq/l but multiplied by M

smp

/ y

mg/l

Same formula as C

smp

= eq/l but multiplied by M

smp

* 1000 / y

mol

Same formula as C

smp

= eq/l but multiplied by Vs * 0.001 / y

mmol

Same formula as C

smp

= eq/l but multiplied by Vs / y

g

Same formula as C

smp

= eq/l but multiplied by M

smp

* Vs * 0.001 / y

mg

Same formula as C

smp

= eq/l but multiplied by M

smp

* Vs / y

µg

Same formula as C

smp

= eq/l but multiplied by M

smp

* Vs * 1000 / y