Speed, Stability, Standard amount – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 182: Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 182

Speed

Sets the speed of the burette used for dosing the increments during a
Monotonic IP or Dynamic IP titration.

When carrying out titrations where the reaction is very slow, decrease
the speed value. For titrations in non aqueous media, where a deposit
is created, the speed can also be decreased.

Enter in:

Method parameters menu (Monotonic IP or Dynamic IP).

Range available:

0.01 ml/min to 50 ml/min.

The maximum speed of a burette is the nominal volume
dispensed in 20 s (for 1 ml, 2 ml, 5 ml, 10 ml and 20 ml burettes)
and in 40 s (for 50 ml burette). In the example of a 10 ml burette,
it means that the speed will stay at 30 ml/min if a value is
entered within the range 30 ml/min - 50 ml/min.

Stability

The stability criterion is used to control the stability of the electrode
signal. Selecting a low value will bring about, accurate but long
measurements.

Enter a stability which is close to the default value (50 mpH/min or
3 mV/min). A zero stability criteria will have no effect on the stability
test performed, the reading will be taken into account when the
Acceptance time is exceeded.

Enter in:

Edit method > Parameters menu
Edit electrode > Calibration parameters menu

Range available:

0 to 99 mpH/min or 0 to 99.9 mV/min

Standard
amount

Amount of standard used to calibrate the titrant.

Enter in:

Edit reagent > Standard menu

Range available:

0.001 to 100000 (unit = Amount unit)

Advertising