Applied signal (ac/dc), Archives data lost - cal. data lost - methods kept, Autochaining – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 40: Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 40

Applied signal
(AC/DC)

Specifies the current type (alternative AC or direct DC) to be sent to
the Pt-Pt socket on the Titration Manager. The AC signal frequency is
1.67 Hz. This option is available if mV(i>0) has been selected for
Measurement in the Edit method or Edit reagent menus.

Enter in:

Method parameters menu
Reagent parameters menu

Archives data
lost - Cal. Data
lost - Methods
kept

Instrument internal failure. Only the method parameters have been
kept.

Autochaining

This option is valid for a Coupled method which is not part of a
sequence.

Autochaining = No

At the end of each method run, you must confirm the result in order to
perform the next method. If a Notification message has been
selected, a message is displayed between each method of the Cou-
pled method.

Autochaining = Yes

At the end of each method run, The methods are chained
automatically in the Coupled method. If a Notification message has
been selected, a message is displayed upon starting the first method
(no message is displayed after).

Refer to "Notification message", page 124

.

Enter in:

Edit method menu (for a Coupled method)