Programming tips

Do not forget to declare electrode(s) while setting up reagent parameters, if calibrated.

Do not forget to declare electrode(s) and reagents(s) when programming your method
parameters

If reagents are calibrated, the same electrode must be used in both sample analysis
method and reagent calibration.

If a Sample Changer is used, do not forget to declare one in the Configuration menu.

If no sun icons appear after the method has been selected, check the following points:

1.

Install reagent(s) for selected method,

see "Install reagent", page 108.

2.

If required, calibrate electrode,

see "Calibrate electrodes", page 55.

3.

Calibrate/enter reagent titre,

see "Calibrate reagents", page 55,

see "Enter titre", page 92.

If Sunny icons appear
These are required to run the selected method.

If the Cloudy icon appears:
The electrode/reagent calibration should be performed within 24 hours.
This is a simple warning, it will not stop you from running the analysis.
If the Stormy icon appears:
Titrant required in the selected method is not installed or has not been calibrated.
Electrode required in the selected method has not been calibrated.
If the Question mark icon appears:
It is a programming error, reagent and/or electrode is/are not defined in the
selected method. Revise the method programming.

When a Stormy or a Question mark icon appears, press 1 “Check” . The Titration
Manager will automatically guide you through the operations necessary to solve
the errors encountered.