GLP-Archives
menu

The GLP-Archives (Good Laboratory Practice) command is available
in each window:

To access:

Sample results: enter Main window and press 4

Reagent calibration results: enter Reagent window and press 4

Electrode calibration results: enter Electrode window and press 3

Figure 20: GLP-Archives for sample results

Storage capacity:

Last 60 sample results.
Last calibration of each reagent.
Last calibration of each electrode.

When the GLP-Archives is full and a new result arrives, the
oldest result stored will be the first one removed.

Ground conflict

Ground conflict: External grounding defined in Setup/Configura-
tion and a metal electrode.

An external grounding is defined for the titration system cell in the
Configuration screen and a metal electrode is used by the method.

When a metal electrode is used, select TIM cell external
Gnd = No in the Configuration menu.

Help

Refer to "Check command", page 59

.

High (result
indicator)

Refer to "Result indicators", page 157

.

The RIGHT and LEFT arrow keys allow you to
move from one screen to the other.
1/4 means the first results out of 4 results saved
for the selected method.