Contrast, Controlled by ttl in, Copy electrode – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 63: Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 63

Contrast

The contrast of the display can be adjusted in the Main window.

press 0 to increase the contrast

press 7 to decrease the contrast

Controlled by
TTL IN

Refer to "Auxiliary input", page 41

.

Copy electrode

This procedure is used to create an electrode, without having to enter
data nor program the corresponding calibration method.

1.

Enter the Electrodes window.

2.

Press 4 then 1.

3.

In the Function field, select the function according to the elec-
trode type then press

,

see "Electrode type", page 89

.

4.

Select the ID field and press

✓ .

5.

Select From = Catalogue.

6.

In the ID field, select an electrode name from the Catalogue list.

7.

In the id field, you can identify the electrode by assigning a
second id name. This electrode will be called "IDid".

8.

Press 1 to confirm then 2 Copy from electrode.

9.

In the Library field, select the Preprogrammed or User list.

10. In the ID field, select the electrode to be copied from the list of

available electrodes.

11. Press 1 to confirm. The electrode is created and saved in the

User list.

If you selected the option Preprogrammed, the list is limited to
electrodes of the same type as the "copied" electrode.

If you selected User, the list is limited to electrodes having the
same function (pH, mV (i=0), mV(i>0), T°C, Reference or
Ground) as the "copied" electrode.

It is not necessary to select Catalogue to create an
electrode using the copy function. An electrode ID created from
Other can also use the copy function.