Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 124

Nb lines per
page (printouts)

When printing page by page (

see "Format (printouts)", page 101

),

this parameter sets the maximum number of lines for one printed
page.

Access:

Setup menu > Configuration

Range available:

25 to 255

No active
electrode
defined in
"method ID"

The method in use, has at least one electrode which has not been
defined. Press

✓ and declare the electrode in the Electrode ID

field of the Method parameters screen.

The reagent in use, uses an electrode which has not been defined for
the reagent calibration. Press

✓ and declare the electrode in the

Parameters screen of the reagent calibration, Electrode - ID
field.

Notification
message

Select Notification = Yes if you want to a message to be
displayed on starting a reagent calibration, a measurement or titration
method.
Type the message (3 lines of 1 to 32 alphanumerical characters).

Enter in:

Edit method menu
Edit reagent menu

Number of
buffers

Available if Calibration request = Yes

Number of pH standards to be used for the calibration. Work with at
least two standards to calculate the electrode sensitivity.
If one standard is only used the zero pH is calculated.

Enter in:

Edit electrode menu

Range available:

1 to 5

Refer to "pH calibration solutions parameters", page 130

.