Working in Routine mode

A.Access methods

A Routine operator has access to all screens for checking purposes

B.Running methods

When working in “ROUTINE” mode, it is necessary to install your titration system according to
the selected method or sequence, prior to running a method or sequence.

1. Select the method or sequence

Refer to "Select method", page 175

.

Refer to "Select sequence", page 175

.

2. Connect the electrodes

Refer to "Electrode connection", page 86

.

3. Check icons

Refer to "Check icons", page 17

.

Depending on the icons displayed, the Titration Manager will automatically guide you through
the steps necessary to run the analysis, see below:

a. Install reagents(s)

Check that the burette is installed, if not,

see "Install burette", page 107

.

Now, install the reagent,

see "Install reagent", page 108

.

b. Calibrate electrode(s)

Now, run the calibration,

see "Calibrate electrodes", page 55

.

c. Calibrate reagent(s) or Enter titre

Now, run the calibration or enter the titre.

Refer to "Reagent calibration", page 145

.

Refer to "Enter titre", page 92

.

d. Run the method or the sequence

Refer to "Running a method (preliminary steps)", page 162

.

Refer to "Running a sequence with a sample changer", page 166

.