Reagent calibration parameters, See "reagent calibration parame, Titra lab – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 147

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 147

Reagent
calibration
parameters

This menu contains the general parameters concerning the reagent
calibration method (identification of the electrodes used in particular).
Titration parameters necessary to run the calibration are also
programmed.

Do not forget to select the electrode and temperature sensor
created in this menu!

To access:

1.

From the Reagent window, press 5.

2.

Select the reagent to be edited.

3.

Press 2 Edit reagent and check that the
Titre = Calibrate option has been selected.

4.

Edit the calibration general parameters.

5.

Use the LEFT/RIGHT arrow keys to move to the last Edit reagent
display.

6.

Press 1 Calibration parameters.