Environmental operating conditions:

Interior use only

Ambient temperature:
- working range: 5 to 40°C
- storage: -20 to 60°C
- transport: -40 to 60°C

Relative humidity:
20 to 80 % with temperature between 5 and 31°C. Above 31°C, the interval amplitude
decreases linearly from 20 - 80% at 31°C to 20 - 50% at 40°C.

Altitude

≤ 2000 m.

Level of pollution : 2

Transitory overvoltage: class II.

International standards

Refer to "International Standards", page 202.

Burette specifications according to ISO 8655-3

Burette

Nominal

Maximum permissible Maximum permissible

stand

volume

systematic errors

random errors

Type

ml

± %

± µl (a)

± % (b)

± µl (c)

B501

< 1

0.6

6

0.1

1

B505

5

0.3

15

0.1

5

B510

10

0.2

20

0.07

7

B525

25

0.2

50

0.07

17.5

B550

50

0.2

100

0.055

25

(a)

Expressed as the deviation of the mean of a tenfold measurement from the nominal volume or from

the selected volume (see ISO 8655-6:—, 8.4).

(b)

Expressed as the coefficient of variation of a tenfold measurement (see ISO 8655-6:—, 8.5).

(c)

Expressed as the repeatability standard deviation of a tenfold measurement (see ISO 8655-6:—, 8.5).