Create electrode, Create method, Titra lab – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 66

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 66

Create electrode

1.

Enter the Electrode window.

2.

Press 4 then 1.

3.

Select the electrode function,

see "Electrode type", page 89

.

4.

Press

✓ in the ID field.

5.

Select From = Other.

The option From = Catalogue allows you to create an
electrode from a list of Radiometer Analytical electrodes.

6.

Enter the electrode name (up to 16 alphanumeric characters).

7.

In the Confirm creation screen:

For pH, mV or ISE electrodes only; select the electrode type.

Refer to "Electrode type", page 89

.

For combined pH or single pH electrodes; enter the internal pH
(pH int) of the electrode.

Refer to "pH int", page 130

.

For combined pH, Metal/Redox or ISE electrodes or for a
Conductivity electrode; select if the electrode has a built-in
temperature sensor or not.

For reference electrodes or combined pH, Metal/Redox, ISE
electrodes; enter the potential of the reference (in mV) versus the
Standard Hydrogen Electrode.

Refer to "Potential versus SHE", page 131

.

8.

Press 1 to create the electrode. The Edit electrode menu is
displayed. Enter the electrode definition parameters.

Create method

1.

Enter the Main window.

2.

Press 5 then 1.

3.

Press

✓ in the New method screen.

4.

Enter a method name (up to 17 alphanumerical characters).

5.

Press 1 to create the method. The Edit method menu is
displayed.

6.

Enter the method parameters.

Refer to "Programming method", page 138

.