Equation formula, Titra lab – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 93

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 93

Equation
formula

One or two user defined equations can be entered to calculate results
using user required units.

The formula specifies how the calculations should be performed. A
formula can use up to 64 characters and can consist of operators and
operands, that are derived from the completed titration data.

Enter in:

Edit method > Result

Table 5: Equation formula: operands

Operands

1.23, -24, -2.34E-5

Constants

pi

3,141592654

e

2,718281828 (base of the Natural loga-
rithm)

Ri where i = 1 to 4

Result no. i calculated for the method

Ei where i = 1 or 2
(for EP) and 1 to 4
(for IP)

Potential difference or pH measured at
EP/IP no. i

Vi where i = 1 or 2
(for EP) and 1 to 4
(for IP)

Cumulate volume delivered at EP/IP no. i

V

Total volume delivered

SA

Sample amount

T

Last temperature measured (in °C)

CT

Titrant concentration

CE

Excess reagent concentration

VE

Excess reagent volume

Bi with i = 1 or 2 (for
EP) i = 1 (for IP)

Blank volume for EP/IP no. i

F

Results factor