TitraLab 840 and 845 Reference Manual

Page 157

Result
indicators

Appear with the result at the end of a run.

OK: accepted by the instrument (result lies within the acceptance
limits).

Low: alarm, electrode calibration rejected by the instrument (result
lies below the acceptance limit).

High: alarm, electrode calibration rejected by the instrument (result
lies above the acceptance limit).

Time max: the measurement has been accepted at the end of the
Acceptance delay.

0IP: no inflection points found.

IP>1: more than one inflection point has been found in the min/max
ordinate window,

see "Min. ordinate Max. ordinate", page 119

.

Result unit

Unit used for the result.

Enter in:

Edit method > Results

Range available:

If Sample unit = ml or µl: mol, mmol, eq/l, meq/l, mol/l, mmol/l, g/l,
mg/l, mg/ml or µg/ml

If Sample unit = g or mg: mol, mmol, eq/kg, meq/kg, mol/kg, mmol/kg,
g/kg, mg/kg, mg/g, µg/g, % or ppm.

The result can be expressed as a mass (g, µg, mg) or a volume (µl or
ml).