Direction

Determines whether the addition of titrant increases or decreases the
pH or mV value.

This parameter informs the Titration Manager if the end point or the
Stop point is already passed before the start of the titration.

Example:

If the titration is performed with an acidic titrant on an alkaline sample,
the addition of titrant will decrease the pH value. In this case, select
Direction = Decreasing pH.

Enter in:

Edit method > Parameters menu
Edit reagent > Calibration parameters menu

Direct
measurements

Refer to "Display measurement", page 73

.

Display contrast

Refer to "Contrast", page 63

.