Volume, Wrong buffer, Wrong type (sac error) – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 195: Zero ph, Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 195

Volume

Refer to "Reagent addition volume", page 145

.

Wrong buffer

While running an electrode calibration, the buffer analysed has not
been identifed by the instrument.

End the calibration cycle. Check the buffer then start a new calibration
cycle.

Wrong type
(SAC error)

Sample changer error: the sample changer connected does not corre-
spond to the one programmed in the Setup > Configuration menu.

Example:
A SAC80 is connected to the Titration Manager, SAC socket and you
have declared a SAC850 in the Setup > Configuration menu.

Review the Setup > Configuration parameters and/or connect the cor-
rect model of Sample Changer to the SAC socket of the Titration
Manager.
Refer to the Sample Changer User’s Guide, part no. D21T085.

Zero pH

This is the pH value (pH

0

) at which the potential is zero. The zero pH

is calculated after a one-point calibration.

Refer to "pH0(25)", page 129

.