TitraLab 840 and 845 Reference Manual

Page 102

Function

Select the electrode function relative to the electrode in use.
The possible electrode functions are:

pH,

mV (i = 0),

mV (i > 0),

T°C,

Reference,

Ground.

Refer to "Electrode type", page 89

.

Fuses

For continued protection replace the fuse with one of a high
interrupting capacity, same type and rating:
2 x fuses, slow blow, 1.0 A (5 x 20 mm), part no. 450-020.
To replace the fuses:

Figure 19: Fuse replacement

hook

3. Remove the fuse holder

4. Replace the used fuses with
ones of the same type and rating

1. Switch off the instrument

5. Put the cap back in place

2. Disconnect line cord