Edit sequence menu, Menu, Titra lab – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 81

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 81

Edit sequence
menu

1.

Perform steps 1 to 6 of the "Programming sequence" procedure.

Refer to "Programming sequence", page 139

.

2.

In the Sequence/Sample stack window (access: press 2
Sequence/Sample stack from the main window),
press 3 Edit sequence.

Select the methods to be included in
the sequence. Use the Add method
(key 1)/ Insert method (key 2)
procedures.

In the title bar, “x/y” (1/1) indicates the
position "x" occupied by the method in
the sequence and "y" the total number
of methods listed in the sequence.

The ID and type of the selected
method cannot be modified at this
level. They are defined in the Add
method or Insert method
menu.

Refer to "Add method menu", page 36

.

Refer to "Insert method menu", page 106

.

3.

At the line Number of samples, enter the number of times a
method must be repeated within the sequence.
At the line Beakers: [F;L], the instrument displays the positions F
and L occupied by the first and last beakers In the sequence.

4.

If a SAC850 is used and declared in the Configuation menu, you
can also choose a sample preparation number used for the
selected method,

see "Sample preparation no.", page 172

.

The last step of the sequence programmation consists of editing the
sample stack,

see "Sample stack", page 174

.