TitraLab 840 and 845 Reference Manual
Page 158
Results
The following results are displayed automatically at the end of a run.
pH calibration
Mean of zero pH (pH0(25))
Mean of sensitivity
Reagent calibration
Titre
Mean ± standard deviation
Measurement method
Measurement
Mean ± standard deviation
Titration method
Based on the volume found at the End point or Inflection point.
Mean and standard deviation for each of the defined result 1 to 4.
In End point, IP monotonic, IP dynamic and IP continuous modes, up
to 4 results and 1 user-defined equations can be calculated.
Results by
difference/
cumulate
Results by difference (ml)
Volumes between two End points or Inflection points are used for
result calculation.
Results by cumulate (ml)
Total volume delivered up to the End point or Inflection point is used
for result calculation.
This setting has no influence on the result calculation for the
first equivalence point detected.
Enter in:
Edit method > Results (End point, IP monotonic, IP dynamic or IP
continuous methods)