Electrode
window

This window contains all the information and operations concerning
the electrodes.

To access:

Use LEFT/RIGHT arrow keys from the Main window.

Electrodes are
different

This message appears if the active electrode used for the reagent
calibration is different from the electrode used in the method.

Check the method and reagent calibration parameters menus.

Empty burette

1.

Enter the Burette functions menu.

2.

Select the burette in the Address field,

see "Address", page 38

.

3.

Press 2 Empty. The burette is emptied at maximum speed
(approximate emptying time: 25 s).

Empty
sequence

Involves removing all the methods present in the sequence.

Proceed as follows:

1.

Select the sequence to be emptied,

see "Select sequence", page

175

.

2.

Press 3 Sequence/Sample stack then 3 Edit sequence.

3.

Press 3 Delete.

4.

Press 2 Delete sequence then

✓ to confirm or press Esc to leave

the screen without emptying the sequence.

To remove a particular method from a sequence,

see "Remove

method from a sequence", page 153

.