Electrode window, Electrodes are different, Empty burette – Hach-Lange TITRALAB 840_845 Reference Manual User Manual
Page 90: Empty sequence, Titra lab

Titra
Lab
®
TitraLab 840 and 845 Reference Manual
Page 90
Electrode
window
This window contains all the information and operations concerning
the electrodes.
To access:
Use LEFT/RIGHT arrow keys from the Main window.
Electrodes are
different
This message appears if the active electrode used for the reagent
calibration is different from the electrode used in the method.
Check the method and reagent calibration parameters menus.
Empty burette
1.
Enter the Burette functions menu.
2.
Select the burette in the Address field,
3.
Press 2 Empty. The burette is emptied at maximum speed
(approximate emptying time: 25 s).
Empty
sequence
Involves removing all the methods present in the sequence.
Proceed as follows:
1.
Select the sequence to be emptied,
.
2.
Press 3 Sequence/Sample stack then 3 Edit sequence.
3.
Press 3 Delete.
4.
Press 2 Delete sequence then
✓ to confirm or press Esc to leave
the screen without emptying the sequence.
To remove a particular method from a sequence,