Equivalent point determination (ip methods), Titra lab – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 97

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 97

Equivalent point
determination
(IP methods)

Incremental addition of the titrant (Monotonic/Dynamic IP
methods):

acquisition of the titration curve E (electrode) = f (volume of titrant
dispensed),

determination of the first derivative curve using a smoothing proc-
ess (depending on the value set for IP filter),

determination of the second derivative curve,

determination of the maxima of the first derivative curve using the
IP filter parameter. These maxima correspond to the zeros of the
second derivative curve.

determination of the equivalent point: for this point the maximum of
the first derivative curve must be equal to or greater than the preset
IP reject method parameter.

Figure 18: Determination of an equivalent point - Monotonic or Dynamic IP

method

An equivalent point is determined and plotted live on the titration
curve. The method is stopped after a given number of equivalent
points (1 to 4) has been detected.

E or pH