Address, Address conflict (between electrode ids), Alarm: locked – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 38: Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 38

Address

The position where the electrode and burette are placed during
operation:

Electrode address

The electrode address is defined using the format “TIM/y” where “TIM”
corresponds to the instrument (TIM840 or TIM845) where the
electrode is connected and “y” corresponds to the socket.
For example TIM/E1, indicates that the electrode is connected to E1
socket on the Titration Manager.

Refer to "Electrode connection", page 86

.

Burette address (Option available with TitraLab 845)

The burette address is defined using the format “TIM/y” where “TIM”
corresponds to the TIM845 Titration Manager where the burette is
placed and “y” corresponds to the position.
For example TIM/1, indicates that the burette is placed on the TIM845
in position 1.

Figure 1: Burette positions

Address
conflict
(between
electrode IDs)

Two electrodes have been defined at the same address.

Enter the Edit electrode menu and modify the address of one of the
electrodes.

Alarm: Locked

The user cannot bypass an electrode calibration if the last result
obtained lies outside the acceptance range.

Enter in:

Setup menu > Access routine mode

Position 2