Remove burette

If no reagent is installed

1.

Enter the Burette functions menu.

2.

Press 5 Remove burette then press

✓. The piston is moved to its

lowest position (filling operation).

3.

Check the burette data, remove the burette and press

✓.

If a reagent is already installed
The reagent is removed and the burette flushed before removal.

1.

Enter the Burette functions menu:

2.

Press 5 Remove burette then press

✓. The piston is moved to its

uppest position (emptying operation).

3.

Remove the reagent bottle and press

✓. The burette is filled then

emptied.

4.

Install a rinse bottle containing demineralised water and press

✓.

The burette is flushed,

see "Flush burette", page 101

.

5.

Remove the rinse bottle and press

✓. The burette is filled,

emptied then filled again. At the end, the piston stops at its lowest
position.

6.

Check the burette data, remove the burette and press

✓.

Remove method
from a
sequence

1.

Press 3 Sequence/Sample stack in the Main window.
If this option is not available, enter the Setup menu > Configura-
tion and declare a SAC80 or SAC850 Sample Changer (Supervi-
sor access only).

2.

Press 3 Edit sequence.

3.

Select the method to be removed using the LEFT/RIGHT arrow
keys.

4.

Press 3 Delete.

5.

Press 1 Remove method then

✓ to confirm or press Esc to leave

the screen without removing the method.

The method is removed from the sequence but is not deleted.
To remove all the methods from a sequence,

see "Empty se-

quence", page 90

.