Communication failure (sac error), Concentration, Concentration unit – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 60: Configuration menu, Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 60

Communication
failure (SAC
error)

The data transmission between the sample changer and the worksta-
tion cannot be performed properly.

Check the cable connections and make sure that the sample changer
is switched on and connected to the workstation via the RS232 serial
cable.

Press the workstation key

✓ or Stop and restart the sequence. It is

not possible to continue the sequence from the point it stopped.

Concentration

Concentration of the titrated species present in the standard.

Enter in:

Edit reagent > Standard

Range available:

0.001 to 10000 (unit = Concentration unit)

Concentration
unit

Standard concentration unit.

Enter in:

Edit reagent > Standard

Range available:

If Amount unit = ml or µl: eq/l, meq/l, mol/l, mmol/l, g/l, mg/l, mg/ml
If Amount unit = g or mg: eq/kg, meq/kg, mol/kg, mmol/kg, mg/g,
mg/kg,%

Configuration
menu

Press Stop 3 seconds in the Main window then press 1.

Contains the configuration parameters for the instrument. .

Refer to "Setup menu", page 178.