Sequence/SAC
sequence

Unlike Coupled methods, sequences also allow you to link reagent
calibration and electrode calibration methods.

A sequence is a chain of up to 3 methods that will be carried out in
different beakers and in a defined order given by the operator.

The beakers are changed automatically using a SAC80 or SAC850
Sample Changer.

Example:

Method 1 performed on 2 samples using 2 test portions
(2 x 2 beakers),

Method 2 performed on 1 sample with 3 test portions (1 x 3 beakers).

Thus 7 beaker system will run as follows:

Figure 32: Sequence of methods

The number of samples is entered in the Edit sequence menu and the
number of test portions in the Edit method screen.

Refer to "Select sequence", page 175

.

Refer to "Programming sequence", page 139

.

Refer to "Running a sequence with a sample changer", page 166

.

Method 1

Method 2

Sample 1

test portion 1/2

test portion 2/2

1

= beaker number 1

2

Sample 2

test portion 1/2

test portion 2/2

3

4

Sample 3

test portion 1/3

test portion 2/3

test portion 3/3

5 6 7

1