Ph calibration results parameters, Ph calibration solutions parameters, Ph int – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 130: Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 130

pH calibration
results
parameters

Refer to "Results menu", page 160

.

pH calibration
solutions
parameters

Refer to "Solution menu", page 181

.

pH int

This parameter is available when creating a single or a combined pH
electrode with the option From = Other.

Refer to "Create electrode", page 66

.

This is the internal pH of a single or combined pH electrode.
The pH int is used for the buffer recognition.
The table below gives the pH int of Radiometer Analytical pH
electrodes:

Table 7: pH int of Radiometer Analytical electrodes

Range available:

0.00 to 14.00 pH

Electrode

pH int

PHG211, PHG301, PHG311, XG100, XG200,
XG250

6.06 pH

PHG201

6.65 pH

PHC2001, PHC2011, PHC2085, PHC2401,
PHC2501, PHC2601, PHC3001, PHC3011,
PHC3081, PHC3185, XC100, XC111, XC120,
XC161

6.65 pH

PHC4000

6.80 pH