Electrode
calibration
stack

The electrode calibration stack defines individual data for each buffer
solution beakers present in an electrode calibration sequence.

1.

In the Configuration menu, declare the model of sample changer
used (SAC80 or SAC850).

Refer to "Configuration menu", page 60

.

Depending on the model of sample changer used, enter the
sample changer configuration parameters.

Refer to "Sample changer", page 169

.

2.

Enter the Electrode window and press 1 Calibrate electrodes.

3.

Select the electrode to calibrate.

4.

Press 2 Calibration sequence.

Label the beakers indicating the running number in the
sequence, for example: 1/5, 2/5 etc.... and the name of the
buffer solution.
Place the beakers in the numbered position on the sample
changer.
If rinses are programmed, position the corresponding rinse
beakers at the right places.

Refer to "Number of static rinses", page 127

.

Refer to "Dynamic rinses", page 77

.

You can print the calibration stack by pressing Print from the
calibration menu.

Press Esc then run the sequence by pressing 1 Run calibration.

<1/5> means the first beaker over 5
programmed in the sequence. Use the LEFT/
RIGHT
arrows to review the other beakers in
the sequence.
The number of beakers is entered in the Edit
electrode menu. The buffer solutions are
selected in the Solutions menu.