Min. speed -
Max. speed
This is the minimum/maximum speed at which the burette will deliver
titrant during an analysis with continuous addition of the titrant (end
point titration for example).
The Min. speed must always be a low value in order to obtain
accurate results. A value of 0.2 ml/min. can be entered for most
applications. A low value entered for the Max. speed will slow
down the titration whereas too high a value will result in an overshoot
of the equivalent point.
Enter in:
Method parameters menu
Reagent parameters menu
Range available:
Min. speed = 0.01 ml/min. to Max. speed
Max. speed = Min. speed to 50.00 ml/min.
Refer to "Burette speed", page 54
Min. Temp. -
Max. Temp.
Acceptance range for the temperature measured in the standard
solution. If the temperature lies outside the defined range the
calibration is stopped.
Enter in:
Edit electrode > Result menu (if Temperature = Probe
or Entered)
Range available:
Min. Temp. = -9°C to + Max. Temp.
Max. Temp = Min. Temp. to +99°C