Burette volume

The burette volume is entered while installing or replacing the burette.
This volume is indicated on the burette.

Enter in:

Install burette or Replace burette menu.

Range available:

1 ml, 5 ml, 10 ml, 25 ml or 50 ml.

Calculate with
EP no.

End point number (1 to Number of EP) for which the volume will
be used to calculate the result. Depending on the option selected for
Resul
ts by cumulate or difference, this will be the volume delivered
from the titration start to the end point no. "n" or the volume delivered
between the end points no. “n-1” and “n”.

Enter in:

Edit method > Results

Calculate with
IP no.

Inflection point number (1 to Number of IP) for which the volume
will be used to calculate the result. Depending on the option selected
for R
esults by cumulate or difference, this will be the volume delivered
from the titration start to inflection point no. "n" or the volume
delivered between the inflection points no. “n-1” and “n”.

Enter in:

Edit method > Results

Calibrate
electrodes

Refer to "Electrode calibration", page 82

.

Refer to "Electrode calibration (SAC sequence)", page 83

.

Calibrate
reagents

Refer to "Reagent calibration", page 145

.

Refer to "Reagent calibration (SAC sequence)", page 146

.

Calibration
delay elapsed

This message appears at titration start. A new electrode or reagent
calibration is required. The delay Periodicity entered in the Edit elec-
trode or Edit reagent screen has elapsed,

see "Periodicity", page

129

.

Press

✓ and perform a calibration.