Select electrode, Select method, Select reagent – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 175: Select sequence, Sensitivity, Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 175

Select electrode

Select an electrode for running calibration purpose

1.

In the Main window, select the method using the electrode.

2.

Press 1 to start a calibration. In that case, the electrode available
will be the one specified in the actual working method or
sequence.

Select an electrode for parameter edition purpose
(Supervisor only)

1.

Press 4 in the Electrode window.

2.

In the ID field, select the electrode from the User list.

Select method

Select a method for running method purpose

1.

Press 2 Select a method in the Main window.

2.

In the ID field, select the method from the User list.

Select a method for parameter edition purpose (Supervisor only)

1.

Press 5. in the Main window.

2.

In the ID field, select the method from the User list.

Select reagent

Select a reagent for running calibration purpose

1.

In the Main window, select the method using the reagent.

2.

From the Reagent window, press 1 to start a calibration. In that
case, the reagents available will be those specified in the actual
working method or sequence.

Select a reagent for parameter edition purpose (Supervisor only)

1.

Press 5 in the Reagent window.

2.

In the ID field, select the reagent from the User list.

Select
sequence

1.

In the Setup > Configuration menu, define the model of sample
changer used.

Refer to "Sample changer", page 169

.

2.

In the Main window, press 3 Sequence/Sample stack.

Sensitivity

Measure of the electrode condition. For ideal electrode chains the
sensitivity is 100%. Generally, however, it is lower. It is expressed as a
percentage of the theoretical slope (59.16 mV/pH) of the curve at
25°C and is determined during a calibration on at least 2 points.