Select electrode, Select method, Select reagent – Hach-Lange TITRALAB 840_845 Reference Manual User Manual
Page 175: Select sequence, Sensitivity, Titra lab

Titra
Lab
®
TitraLab 840 and 845 Reference Manual
Page 175
Select electrode
Select an electrode for running calibration purpose
1.
In the Main window, select the method using the electrode.
2.
Press 1 to start a calibration. In that case, the electrode available
will be the one specified in the actual working method or
sequence.
Select an electrode for parameter edition purpose
(Supervisor only)
1.
Press 4 in the Electrode window.
2.
In the ID field, select the electrode from the User list.
Select method
Select a method for running method purpose
1.
Press 2 Select a method in the Main window.
2.
In the ID field, select the method from the User list.
Select a method for parameter edition purpose (Supervisor only)
1.
Press 5. in the Main window.
2.
In the ID field, select the method from the User list.
Select reagent
Select a reagent for running calibration purpose
1.
In the Main window, select the method using the reagent.
2.
From the Reagent window, press 1 to start a calibration. In that
case, the reagents available will be those specified in the actual
working method or sequence.
Select a reagent for parameter edition purpose (Supervisor only)
1.
Press 5 in the Reagent window.
2.
In the ID field, select the reagent from the User list.
Select
sequence
1.
In the Setup > Configuration menu, define the model of sample
changer used.
Refer to "Sample changer", page 169
.
2.
In the Main window, press 3 Sequence/Sample stack.
Sensitivity
Measure of the electrode condition. For ideal electrode chains the
sensitivity is 100%. Generally, however, it is lower. It is expressed as a
percentage of the theoretical slope (59.16 mV/pH) of the curve at
25°C and is determined during a calibration on at least 2 points.