Install reagent, Titra lab – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 108

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 108

Install reagent

Proceed as follows to install a reagent:

1.

Select the method or the sequence which uses the reagent.

2.

Enter the Reagents window. Press 3 to check that the burette is
installed. If not, install the burette by pressing 4,

see "Install

burette", page 107

.

3.

From the Reagents window, press 2 Install reagents.
The identification data of the reagent being installed are
displayed. Press 1.

If a reagent used by the method is already installed at the
address indicated. The instrument will automatically detect the
reagent. All you have to do is to check the ID, address and
position of the delivery tip. Press 1 to end the procedure.

4.

Press 1 to rinse the burette before installing the reagent. Press 2,
if no rinse required.

5.

Install the new reagent bottle, press

✓. The instrument runs a

Flush using the new reagent then fills the burette. The installation
is complete.