Read me, Titra lab, Working in supervisor mode – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 16

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 16

Read me!

An important feature of this Titration Manager interface is that it controls the presence of differ-
ent elements necessary to run the defined application for a selected method/sequence, before
the method/sequence is run.

Working in Supervisor mode

A Supervisor has access to all the libraries for creation purposes.
When programming the Titration Manager in “SUPERVISOR” mode, it is recommended to
work in stages. These stages must be carried out in the order described below:

A.To program method

1.Define your electrode(s)

Identify electrodes (including temperature sensors) to be used for the analysis:
Electrodes can be created from the following lists:
Catalogue,

see "Catalogue list", page 56

.

Other,

see "Others list", page 128

.

Copy from,

see "Copy electrode", page 63

.

When creating the electrode, define if electrode calibration is required (or not), if yes specify
the "periodicity" of the calibrations and the pH standards to be used.

Refer to "Calibrate elec-

trodes", page 55

.

2. Define reagent

Identify reagents to be used for the analysis
Reagents can be created from the following lists:
Catalogue,

see "Catalogue list", page 56

.

Other,

see "Others list", page 128

.

Copy from,

see "Copy reagent", page 64

.

When creating the reagent, define if reagent calibration is required (or titre entered manually),
if yes specify the "periodicity" of the calibrations and the calibration method.

Refer to "Reagent

calibration", page 145

.

If a sample changer is to be used, define the sample changer in the Configuration menu before
programming a SAC sequence.

If you are to perform a calibration, make sure that the electrode(s) used for the calibra-
tion are the same as those used in the method.

Advertising