Beaker menu, Titra lab – Hach-Lange TITRALAB 840_845 Reference Manual User Manual
Page 52

Titra
Lab
®
TitraLab 840 and 845 Reference Manual
Page 52
Beaker menu
Use this menu to prepare a sample or calibration stack. This menu
defines individual data for all the samples or standards used in the
sequence.
Figure 9: Beaker menu (for a sample stack)
To access (for a sample stack):
1.
Press 3 Sequence/Sample stack in the Main window.
A Sample Changer must be declared in the Configuration
menu.
Refer to "Sample changer", page 169
2.
Press 1 Sample stack.
The sequence must have been edited in the Edit sequence
menu.
Refer to "Edit sequence menu", page 81
Refer to "Sample stack", page 174
.
To access (for a reagent calibration stack):
1.
In the Reagent window, press 1 Calibrate/Enter titre.
2.
Press 3 Calibration sequence.
A Sample Changer must be declared in the Configuration
menu.
Refer to "Sample changer", page 169
The reagent calibration method must have been edited
beforehand.
Refer to "Edit reagent menu", page 80
.
Refer to "Reagent calibration stack", page 148
.
To access (for an electrode calibration stack):
1.
In the Electrode window, press 1 Calibrate electrodes.
2.
Press 2 Calibration sequence.
A Sample Changer must be declared in the Configuration
menu.
Refer to "Sample changer", page 169
The electrode calibration method must have been edited
beforehand.
Refer to "Edit electrode menu", page 78
.
Refer to "Electrode calibration stack", page 85
.