Beakers: [f;l, Beep, Blank (yes/no) – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 53: Blank not required, Blank on inflection no, Blank required, Titra lab

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 53

Beakers: [F;L]

The beakers information is displayed in the Edit sequence menu of a
sequence.

It indicates the First and Last positions occupied by the beakers in the
sequence.

Beep

If Yes has been selected, three beeps will sound when a result is
obtained.

Enter in:

Setup menu > Configuration

Blank (Yes/No)

Select Yes to subtract the volume of reagent used to titrate the solvent
from the volume found for the dissolved or diluted sample.

When you run for the first time a method with a blank, you start with
the determination of the blank. The blank volume is saved with the
method. If you start again this method, you have the choice between
running an analysis with the blank alone or running an analysis of the
sample with subtraction of the blank volume to the result.

Enter in:

Edit method menu

Blank not
required

This message appears at the start of a sequence, if a method
sequence programmed with a blank titration has now been
reprogrammed without a blank.

Go to Sequence/Sample stack, Edit sequence menu and remove the
blank titration method from the sequence.

Blank on
inflection no.

Inflection point number (1 to Number of IP) on which the blank volume
will be subtracted.

Enter in:

Edit method > Results (for Monotonic IP, Dynamic IP or Continuous IP
method).

Blank required

This message appears at the start of a titration when a blank has
been programmed in the method, but a blank value is not available.

Press

✓ and run blank.

Advertising