Electrode connection, Titra lab – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 86

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 86

Electrode
connection

1.

In the Main window, press 2 Select method and select the method
to be run. The electrodes to be connected are listed with their
addresses.

2.

Install electrodes on the Titration Manager (TIM), or sample
changer (SAC80 or SAC850).

3.

Connect electrodes to the rear panel socket of the Titration
Manager (TIM). See figure and table below. Example: pHC2401-8
to E1 socket (address TIM/E1).

Refer to "Address", page 38

..

Figure 15: Electrode sockets

Table 3: Connecting electrodes

When installing/checking an electrode system, do not forget to
connect all the electrodes to the Titration Manager and disconnect all
the electrodes that are not used by the method or sequence.
In particular, if you are about to run a pH (or potential at zero current)
measurement, check that there is no double metal electrode con-
nected to the Pt-Pt instrument socket before running the method (dis-
connect this electrode, if any).

P

t

-P

t

R

ef

E1

E2

T

emp

GND

Socket

Electrode

REF

Single reference

TEMP

Temperature

GND

Single metal for cell grounding only

Pt-Pt

Double metal

E1

Others