Sample preparation no, Titra lab – Hach-Lange TITRALAB 840_845 Reference Manual User Manual

Page 172

Advertising
background image

Titra

Lab

®

TitraLab 840 and 845 Reference Manual

Page 172

Sample
preparation no.

The Sample preparation routine gathers all the information required
by a SAC850 Sample Changer to prepare each sample beakers of the
sequence (sample dilution by adding a solvent, stirring delay, sample
degassing, set the sample to a given temperature, start a dynamic
rinse, etc.).
One Sample preparation routine is defined per method executed.
A SAC850 can run up to 10 different sample preparations in a given
sequence.
Sample preparation routines are created and edited from a PC and
TitraMaster 85 PC software. They are loaded in the Titration Manager
memory on connecting the workstation to TitraMaster 85.
On running a sequence, the Titration Manager sends to the sample
changer all sample preparations to be runned for the sequence (one
preparation per method).

Enter in:

Edit sequence menu (with a SAC850)

Range available:

0 to 10
0 means no sample preparation selected.

Advertising