PC keyboard

Specify the PC keyboard in use. For example, English (US) for a
Qwerty keyboard.

Enter in:

Setup menu > Configuration

Range available:

English (US), French, German, Spanish, Italian, Danish, Swedish.

Periodicity

Available if Electrode calibration request = Yes or Titre = Calibrate

Maximum period of time between two calibrations. If the period of time
is exceeded, measurements can no longer be performed using this
electrode or reagent. A new calibration is required except if
Demand : Unlocked has been set in the SETUP > Access Rou-
tine mode menu.

Enter in:

Edit electrode menu
Edit reagent menu

Range available:

1 to 999 days

pH0(25)

This is the pH of the solution at 25°C at which the measured electrode
potential is equal to zero. The potential developed depends on both
electrodes; the reference and the glass. These potentials may vary to
bring about an electrode drift. This drift can be compensated by
frequent calibrations.

Normally the following ranges are used by default (6.850 pH and
7.200 pH)

Enter in:

Edit electrode > Results

Range available:

pH(0)25 min. = -9.000 pH to pH0(25) max.
pH0(25) max. = pH0(25) min. to 23.000 pH

pH buffer

Refer to "Solution menu", page 181

.